商品介绍
DescriptionQ & AManualS.D.S
Lactate Assay Kit-WST can quantify lactate (the lower limit: 0.02 mmol/l), one of the glycolysis metabolites. Because the Lactate Assay Kit-WST corresponds to 96 well micro-plates, it allows for multiple samples to be measured at once. Lactate quantification has been widely used as a marker for monitoring changes in intracellular metabolic pathway in many different areas of cancer research. In recent years, there have been an increasing number of reports on using lactate as an indicator of stem cell differentiation or as an indicator in diabetic research.


PrincipleThe kit is used to detect lactate in cell culture medium by measuring the absorbance of WST formazan produced according to quantities of lactate. The kit includes a “Lactate Standard” which can be used to quantify the concentration of lactate found within samples by creating standard curve.


Simple Assay Procedure


Stability

Dojindo’s stock solution is stable for 4 months under refrigerator. Your assay can be done anytime without the need freeze the product and wait for defrosting.


Standard curveStandard curve can be prepared by using Lactate Standard included in the kit. The concentration of lactate can be measured. The concentration of lactate can be evaluated by diluting the samples if the concentration is over 1 mmol/l.

Can I quantify D-Lactate levels?
No, the kit is used for the quantification of L-Lactate levels.
How stable is the Working solution?
You can’t store the Working solution. Please prepare the Working solution prior to use. Please protect from light because the Working solution is light-sensitive. The Working solution is stable for 4 hours at room temperature with protection from light.Once the Working solution is exposed to light, the color of the solution becomes from red to orange. It caused to increase background.
Can I measure samples that contain a reducing agent?
You can’t measure an accurate Lactate level because WST dye interferes with the reducing agent. If you add test material which possesses reducing properties, please prepare a few wells that only contain the medium without cells and only test material for background control. Please subtract the background from the absorbance of standard curve or samples
Are there protocols for measuring intracellular lactate level?
(1) Collect cells (1 × 105 cells * 1) in a microtube.(2) Centrifuge at 300 × g for 2 minutes and remove supernatant.(3) Add 300 μl of cold PBS and suspend by pipetting. Centrifuge at 300 × g for 2 minutes and remove supernatant.(4) Add 300 μl of 0.1% Triton-X * 2 and vortex for 1 minutes to prepare for cell lysate.(5) Centrifuge at 8000 × g for 5 minutes and collect supernatant.(6) Deproteinize the solution prepared at step (5) with an ultrafiltration membrane filter (fraction molecular weight: 10 K) * 3 to obtain a measurement sample.

* 1 For HeLa cells, more than 1 × 105 cells are necessary to detect lactate level over 0.02 mmol / l.* 2 When SDS is included in the cell Lysate, color development can be inhibited. Please avoid using buffer containing SDS.* 3 Since endogenous Lactate dehydrogenase (LDH) causes background please remove LDH by deproteinization treatment.

How many samples can I measure?

* The number of samples that can be recorded when the standard curve and sample is measured in triplicates.

*A plate layout for Lactate standard solution and sample

I do not have a 450 nm filter. What other filters can I use?
You can use filters with an absorbance between 450 nm and 490 nm. However, the absorbance value can become lower if the sample is not measured at 450 nm.
Our samples did not change in color, are there any reasons for this?
Lactate level contained in the sample may be lower than the detection limit (0.02mmol/l) that can be measured using this kit. If the lactate level is below 0.02mmol/l, please consider other measurement method such as LC-MS.If you dilute the sample, lactate level contained in the diluted sample may be lower than 0.02mmol/l.Please lower the dilution ratio and adjust the concentration of the sample over the detection limit that was measured using this kit.

Related Categories Cell Proliferation / Cytotoxicity Intracellular Fluorescent Probes Cellular Metabolism

品牌介绍
Dojindo细胞分析细胞活力和细胞毒性测定用于药物筛选和化学物质的细胞毒性测试。Dojindo开发了高度水溶性的四唑盐,称为WST。WST-8是高度稳定的WST,用于Cell Counting Kit-8(CCK-8)。由于WST-8甲maz是水溶性的,因此不会形成晶体。因此,不需要诸如MTT测定的增溶过程。此外,CCK-8的检测灵敏度高于其他四唑盐,例如MTT,XTT,MTS或WST-1。WST检测机制 ß-半乳糖苷酶检测试剂细胞增殖/细胞毒性转染细胞染色细胞内荧光探针细菌染色微生物活力测定干细胞分化SPiDER-ßGal线粒体检测细胞代谢应用产品展示细胞生长检测,药物筛选,比色/荧光检测细胞计数试剂盒-8细胞计数试剂盒8 + 96孔有机硅定向剂细胞计数试剂盒-F细胞毒性LDH检测试剂盒-WST 96孔有机硅定向剂MTT了解检测机制的差异:点击这里细胞周期分析细胞周期测定溶液深红色细胞周期测定溶液蓝色